Deactivating Or Activating Flow Monitoring - Dräger Savina 300 Instructions For Use Manual

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

If calibration was not successful
If calibration was not successful, the message
Flow calibration failed. is displayed. The
expiratory portion of the flow waveform and the
measured values VT
, MV, and PEEP are not
e
displayed.
 Replace flow sensor.
Savina 300 automatically calibrates the new flow
sensor.
Deactivating or activating flow
monitoring
Flow monitoring can be deactivated, e.g.:
– If the flow sensor has failed and cannot
currently be replaced.
– To permit ventilation in the event of major tube
leakage.
Savina 300 cannot determine the following
measured values when flow monitoring is
deactivated:
– MV
– MV
spon
– MV
leak
– VT
e
– VT
spon
– R
– C
– RSB
– PEEP
Instead of the measured value, the following is
displayed in the parameter field and on page
Trends/Data > Measurements: Off
Expiratory flow monitoring cannot be fully
substituted via replacement monitoring. Set the
minute volume alarm limits of the replacement
monitoring accordingly.
Instructions for use Savina 300 SW 5.n
WARNING
Risk of patient injury
If flow monitoring is deactivated, Savina 300
cannot monitor the patient completely.
Ensure that appropriate replacement monitor-
ing is available immediately.
WARNING
Risk of patient injury
No apnea monitoring takes place when flow
monitoring is deactivated.
Use independent apnea monitoring.
Deactivating flow monitoring
1 Touch the Sensors... button in the main menu
bar. The Flow (A) page appears by default.
Sensors
A
B
C
2 Touch the Off button (C) and confirm with the
rotary knob.
The measured values are no longer displayed. The
alarm function is deactivated. The following
information is displayed in the header bar: Use
external flow monitoring.
Activating flow monitoring
Reactivate flow monitoring after replacing the flow
sensor or as soon as possible.
 Touch the On button (B) and confirm with the
rotary knob.
Monitoring