Dräger Savina 300 Instructions For Use Manual page 153

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Priority
Alarm
!!!
700
Device failure
!!
310
Device temperature high
intPEEP not confirmed
!
291
P
!
291
supp
!!
083
etCO
high
2
!!
082
etCO
low
2
!
060
Expiration hold
interrupted
Instructions for use Savina 300 SW 5.n
Cause
The device failed.
The internal device tem-
perature is too high.
The setting was changed,
but the change was not con-
firmed.
not confirmed
The setting was changed,
but the change was not con-
firmed.
Upper alarm limit for end-ex-
piratory CO
has been exceeded.
Lower alarm limit for end-ex-
piratory CO
has been exceeded.
The "Exp. hold" button was
pressed too long.
Remedy
Disconnect patient from the
device and continue ventila-
tion without delay using an-
other independent ventilator.
Check dust filter for soiling
and replace if necessary.
Check the setting.
If necessary, adjust the set-
ting and confirm with the
rotary knob.
Acknowledge message by
pressing "Alarm reset" key.
Check the setting.
If necessary, adjust the set-
ting and confirm with the ro-
tary knob.
Acknowledge message by
pressing "Alarm reset" key.
Check patient condition.
concentration
2
Check ventilation settings.
Adjust alarm limit if neces-
sary.
Perform CO
tion if necessary.
Check whether the cuvette
windows are soiled.
Check patient condition.
concentration
2
Check ventilation settings.
Adjust alarm limit if neces-
sary.
Perform CO
tion if necessary.
Check whether the cuvette
windows are soiled.
Release "Exp. hold" button.
Troubleshooting
zero calibra-
2
zero calibra-
2
153