Dräger Savina 300 Instructions For Use Manual page 161

Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Priority
Alarm
!
120
MEDIBUS communica-
tion failed
!!
391
Medium-priority test alarm
!!
110
Microfilter blocked
!!!
320
Microfilter is missing
!!!
180
MV high
!!!
190
MV low
Instructions for use Savina 300 SW 5.n
Cause
MEDIBUS communication
failed.
Alarm signals are being test-
ed.
Microfilter is extremely
soiled.
Microfilter missing or incor-
rectly inserted.
The minute volume exceeds
the upper alarm limit.
Water in flow sensor.
The minute volume has fall-
en below the lower alarm
limit.
Obstruction
Leakage or disconnection.
Device failure
Troubleshooting
Remedy
Ventilation functions are not
affected. Check MEDIBUS
connection. Check MEDIB-
US settings.
Complete or cancel device
check.
Replace microfilter.
Insert microfilter.
Check patient condition.
Check ventilation settings.
Adjust alarm limit if neces-
sary.
Drain water trap of breathing
circuit. Dry flow sensor.
Check patient condition.
Check ventilation settings.
Adjust alarm limit if neces-
sary.
Check patient condition.
Check breathing circuit.
Check tube or mask.
Check breathing circuit for
tight connections.
Make sure that the
expiratory valve is properly
engaged.
Make sure that the tube or
mask is connected correctly.
Disconnect patient from the
device and continue ventila-
tion without delay using an-
other independent ventilator.
Contact specialized service
personnel.
161