Operation
4 Remove the corrugated hose of the breathing
circuit (G) from the inspiratory port of the
Y-piece and connect it to the catheter
connector (D).
5 Connect the other end of the corrugated
hose (F) to the inspiratory port of the Y-piece.
Connecting the nebulizer hose
Connect the nebulizer hose (I) onto the
nebulizer port (H).
Performing pneumatic medication
nebulization
Prerequisites:
– Medication nebulizer is filled in accordance with
the corresponding instructions for use.
– The correct functioning of the medication
nebulizer is ensured.
CAUTION
Insufficient medication nebulization
A medication nebulizer malfunction is not
detected by Savina 300.
Check the correct functioning of the medication
nebulizer. Check whether aerosol is generated.
94
Switching on medication nebulization
Press the
Savina 300 starts nebulization. The nebulization
time is 30 minutes. The
remaining time are displayed in the header bar.
Terminating medication nebulization
prematurely
Press the
After medication nebulization
Savina 300 automatically switches off the
medication nebulizer after the nebulization time has
elapsed.
H
Savina 300 automatically cleans the flow sensor by
heating and performs calibration following
medication nebulization.
I
1 Remove any residual medication. Observe the
instructions for use of the medication nebulizer.
2 If a bacterial filter is used to protect the
expiratory valve, replace or remove the
bacterial filter.
Performing medication nebulization with
the Aeroneb Pro nebulizer
Observe the instructions for use of the Aeroneb
Pro nebulizer.
Observe the section "Information on breathing
circuits and additional components"
on page 55.
Observe the safety information on medication
nebulization, see page 92.
NOTE
Do not switch on medication nebulization on
Savina 300 as the Aeroneb Pro nebulizer does
not require a nebulizer flow from Savina 300.
Instructions for use Savina 300 SW 5.n
Nebul. on/off key.
symbol and the
Nebul. on/off key.