Brainlab CRANIAL/ENT OPTICAL User Manual page 55

Instrument
Table of Contents

Advertisement

POINTERS
RF Emissions Interferences
Softouch and Z-touch use RF energy only for its internal function. Therefore, the RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
Softouch and Z-touch are suitable for use in all establishments other than domestic and those
directly connected to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Standards
CISPR 11
IEC 61000-3-2
IEC 61000-3-3
NOTE: U
Separation Distances
Softouch and Z-touch are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of Softouch or Z-touch can help
prevent electromagnetic interference.
This is accomplished by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and Softouch or Z-touch, according to the maximum
output power of the communications equipment.
Rated Maximum Output
Power of Transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, d can be determined using
the provided equations.
P = maximum output power rating of the transmitter in watts (W)
d = recommended separation distance in metres (m)
Instrument User Guide Rev. 2.3 Cranial/ENT Optical
Emission Test
RF emissions
Harmonic emissions
Voltage fluctuations/flicker emissions
is the AC main power voltage prior to the application of the test level.
T
Separation Distance According to Frequency of Transmitter
(m)
150 kHz to 800 MHz 80 MHz to 800 MHz
0.12
0.37
1.17
3.69
11.67
Compliance
Group 1, Class B
N/A
800 MHz to 2.5 GHz
0.23
0.73
2.30
7.30
23.00
b
55