Dräger Narkomed 3 Technical & Service Manual page 80

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

REPLACEMENT PROCEDURES (continued)
4.15
Convenience Outlet AC Power Filter
The convenience outlet AC power filter is located in the back of the ventilator
box near the AC convenience outlets. Access to the power filter requires lifting
the monitor box. Figure 4-15 shows the location of the filter, its connections
and its mounting arrangement.
4.15.1
Set the System Power switch to STANDBY and remove AC power
from the machine.
4.15.2
Remove the screws holding the monitor box to the ventilator box.
WARNING:
4.15.3
Raise the rear of the monitor box chassis approximately two inches,
and prop the box open.
WARNING:
4.15.4
Remove the two screws holding the safety cover over the AC power
filter, and remove the cover.
4.15.5
Disconnect the AC wiring from each side of the filter. Note the
position of the wires so they can be re-installed in the same manner.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Two people are required to perform the remaining
portion of this procedure.
Ensure that AC power is removed from the
machine before raising the monitor box. Failure to
observe this precaution may cause injury by
electric shock.
4-50
NM3