Dräger Narkomed 3 Technical & Service Manual page 55

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
4.7.12
Perform the following test:
4.7.12.1
4.7.12.2
4.7.12.3
4.7.12.4
4.7.12.5
4.7.12.6
4.7.12.7
4.7.13
Replace the rear cover and its retaining screws.
4.7.14
Connect the pipeline hoses.
4.7.15
Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued)
Open the cylinder valves.
Set the System Power switch to ON.
Set the oxygen flow to five liters per minute.
Set the other gas flow to five liters per minute.
Close the oxygen cylinder valve.
As the oxygen flow decreases, the other gas flow should
also decrease proportionally.
Set the System Power switch to STANDBY.
4-25