ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
5.4
Oxygen Monitor Calibration
5.4.1
Offset Adjustment
5.4.1.1
5.4.1.2
5.4.1.3
5.4.1.4
5.4.1.5
5.4.1.6
5.4.1.7
5.4.1.8
5.4.1.9
5.4.1.10
5.4.1.11
5.4.1.12
5.4.1.13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Set the System Power switch to STANDBY.
Press and hold the
monitor panel, and turn the System Power switch to ON.
The display window should indicate
If the display indicates
required.
If the display does not indicate
adjustment procedure.
Turn the System Power switch to STANDBY.
Disconnect any external serial interface cables from the
back of the monitor housing.
Remove the two monitor housing back cover retainer
screws, and swing open the back cover.
Remove the top shelf from the monitor housing and
disconnect its ground wire.
Repeat Step 5.4.1.2.
Adjust R17 on the top of the monitor (see Figure 5-4) to
100
bring the reading to
Place a small drop of red Glyptal insulating varnish on
the adjustment screw to lock it in place.
Turn the System Power Switch to STANDBY and remove
AC power from the machine.
Reinstall the top shelf on the monitor housing and re-
connect its ground wire.
Close the monitor housing back cover and replace the
retaining screws. Re-connect any external cables that
were previously removed from the serial ports.