REPLACEMENT PROCEDURES (continued)
*
4.10.24.23 Slowly open the N
*
4.10.24.24 The yellow O
4.10.24.25 Adjust the oxygen flow to 2 liters per minute.
4.10.24.26 Slowly open the N
*
4.10.24.27 The yellow O
4.10.24.28 Close the oxygen flow control valve. The nitrous oxide
4.10.24.29 The nitrous oxide flow should stop when the flow of
*
4.10.24.30 If the O
*
4.10.24.31 If the O
4.10.24.32 Close all cylinder valves except the O
4.10.24.33 Set the oxygen flow to 5 liters per min.
4.10.24.34 Open the other gas flow control valves to drain pressure
4.10.24.35 Close the O
4.10.24.36 Close the N
4.10.24.37 Reconnect the pipeline hoses.
4.10.24.38 Replace the rear cover of the vapor box.
4.10.24.39 Perform the PMS Procedure given in Section 6.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
O flow control valve.
2
/N
O FLOW RATIO lamp on the alarm
2
2
panel should light as soon as the ORMC limits the flow
of nitrous oxide.
O flow control valve.
2
/N
O FLOW RATIO lamp on the alarm
2
2
panel should light as soon as the ORMC limits the flow
of nitrous oxide.
flow should decrease proportionally, and the oxygen
concentration should remain between 21% and 29%
oxygen.
oxygen is between 200 and 300 ml per minute.
/N
O FLOW RATIO lamp turns off when the
2
2
nitrous oxide flow drops to between 100 and 200 ml per
minute, proceed to Step 4.10.24.32.
/N
O FLOW RATIO lamp does not turn off when
2
2
the nitrous oxide flow drops to between 100 and 200 ml
per minute, adjust the MPL switch until the O
FLOW RATIO lamp turns off.
from the system.
cylinder valve and the O
2
Press the O
Flush valve to drain oxygen pressure from
2
the system.
O flow control valve.
2
4-36
NM3
/N
O
2
2
valve.
2
flow control valve.
2