PMC PROCEDURE (continued)
( )
6.18.4
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
"F" Bellows Test
6.18.4.1
Set the FREQUENCY to 10.
6.18.4.2
Set the I:E RATIO to 1:2.
6.18.4.3
Adjust the O
6.18.4.4
Adjust the INSPIRATORY FLOW to MED.
6.18.4.5
Adjust the Tidal Volume to 200 ml.
6.18.4.6
What is the Tidal Volume on the Spiromed(B)? ___ ml
(125-250)
6.18.4.7
this step intentionally left blank
6.18.4.8
this step intentionally left blank
6.18.4.9
Adjust the Tidal Volume to 1000 ml.
6.18.4.10
What is the Tidal Volume on the Spiromed(B)? ___ ml
(900-1100)
6.18.4.11
Adjust the INSPIRATORY FLOW to HIGH.
6.18.4.12
this step intentionally left blank
6.18.4.13
this step intentionally left blank
6.18.4.14
Adjust the O
6.18.4.15
this step intentionally left blank.
6.18.4.16
Adjust the Tidal Volume to maximum.
6.18.4.17
What is the Tidal Volume on the test volumeter? ___
ml (>1400)
6-28