Dräger Narkomed 3 Technical & Service Manual page 242

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

PMC PROCEDURE (continued)
( )
6.31
Oxygen Ratio Monitor Control Test
6.31.1
6.31.2
6.31.3
6.31.4
6.31.5
6.31.6
6.31.7
6.31.8
6.31.9
6.31.10
6.31.11
6.31.12
6.31.13
6.31.14
6.31.15
6.31.16
6.31.17
6.31.18
6.31.19
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Open the oxygen and nitrous oxide cylinder valves.
Depress the "O
FLUSH" for 15 seconds.
2
Turn the N
O flow control valve counter-clockwise to its stop
2
position.
Slowly increase the oxygen flow until nitrous oxide begins to flow.
What is the oxygen flow rate? ___(200-300 ml/min.)
Slowly increase the oxygen flow until the "O
LED on the alarm panel is lighted.
What is the nitrous oxide flow rate? ___(150-300 ml/min.)
Set the oxygen flow to 1000 ml/min.
Adjust the nitrous oxide flow until the ORMc controls the nitrous
oxide.
What is the oxygen concentration? ___ % (21-29)
Does the "O
/N
O FLOW RATIO" LED turn "ON" on the ALARM
2
2
PANEL? ___ (Y)
Does the "O
/N
O LOW" message appear in the ADVISORY column?
2
2
___ (Y)
Open the nitrous oxide flow control valve to the stop position.
Adjust the oxygen flow to 1.5 l/min.
What is the oxygen concentration? ___ % (21-29)
Adjust the oxygen flow to 2 l/min.
What is the oxygen concentration? ___ % (21-29)
Adjust the oxygen flow to 3 l/min.
What is the oxygen concentration? ___ % (21-29)
6-38
RETURN TO CD-ROM TABLE OF CONTENTS
/N
2
NM3
O FLOW RATIO"
2
Rev. V