PMC PROCEDURE (continued)
6.39
Multispec (if applicable)
6.39.1
NOTE:
NOTE:
6.39.2
( )
( )
6.39.3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Intentionally left blank
Before performing the accuracy test, the Multispec must be in the
full accuracy mode. The CO
appear on the advisory display.
Before performing the accuracy test, the oxygen analyzer must
display 21% FIO
. The sensor must be exposed to ambient air
2
during the entire Multispec accuracy testing.
Sample Flow Test
6.39.2.1
Verify that the neonatal mode is disabled.
6.39.2.2
Verify that an oxygen supply is connected to the
machine.
6.39.2.3
Examine the sample circuit consisting of the airway
adapter/filter, sample line, and auxiliary water trap.
Replace any components as needed.
6.39.2.4
Connect a capnomed flowmeter test stand to the
patient sample exhaust port.
6.39.2.5
Verify that the sample flow is within 175 to 225
ml/min.
Line Block Test
6.39.3.1
Remove the calibration gas cylinder from the multispec
calibration adapter.
6.39.3.2
Set both flow control valves on the multispec
calibration adapter fully counter-clockwise.
6.39.3.3
Attach the patient sample line to the luer lock fitting
on the calibration adapter.
6-54
RETURN TO CD-ROM TABLE OF CONTENTS
and Agent warm-up messages shall not
2
NM3
Rev. V