ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
5.6A
Multispec 4610 Analyzer Calibration and Flowrate Adjustment
NOTE:
Perform the flow rate adjustment and the oxygen sensor ambient air
calibration before proceeding with the multispec span calibration.
Flowrate Adjustment:
5.6.1.A Connect a test flowmeter to the exhaust port on the sensor interface
panel of the Narkomed 3.
5.6.2.A Verify that a water trap and sample line are connected at the patient
interface panel.
5.6.3.A Turn the System Power switch to ON while simultaneously pressing
the
display window will show
NOTE: From a cold start (AC first applied) it may require a two minute wait
until the pump turns on.
During the first two minutes of operation the AGENT SELECT key
will be disabled, and there will be no line block detection.
During the first eight minutes of operation the ALARMS DISABLE
and CAL keys will be disabled. The fan and the Andros continue
running when the machine is in standby and AC is applied.
The fan and Andros will continue running when the machine is in
STANDBY, and AC is applied.
DISPLAY
WINDOW
%
mmHg
Figure 5-6A: MULTISPEC MONITOR PANEL
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
and
keys on the multispec monitor. The multispec
ET CO2
ET AGENT
INSP AGENT
APNEA
ET CO2 HI
HI
ET CO2 LO
AGENT HI
5-23A
RETURN TO CD-ROM TABLE OF CONTENTS
02
.
CO2/AGT
AGENT
CAL
ALARMS
SELECT
DISABLE
ALARMS
LO
NM3
DISPLAY KEY
SP12112
Rev. B