Dräger Narkomed 3 Technical & Service Manual page 259

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
( )
6.38
Vapor Exclusion System (and Indicator Lamps if applicable)
6.38.1
6.38.2
*
6.38.3
6.38.4
6.38.5
6.38.6
6.38.7
*
6.38.8
6.38.9
6.38.10
6.38.11
6.38.12
*
6.38.13
6.38.14
6.38.15
6.38.16
* Denotes Agent Order Indicators test.
Rev. V
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Set all vapors to zero (0).
Adjust the handwheel on the left position vapor to any
concentration above zero (0).
Verify that the yellow indicator lamp above the left position
vaporizer is illuminated. Press AGENT SELECT on the multispec
monitor and verify the correct blinking agent initial in the monitor
window.
Is it possible to adjust the middle position vapor? ___ (N)
Is it possible to adjust the right position vapor? ___ (N)
Adjust the handwheel on the left position vapor to zero (0).
Adjust the handwheel on the middle position vapor to any
concentration above zero (0).
Verify that the yellow indicator lamp above the center vaporizer is
illuminated. Press AGENT SELECT on the multispec monitor and
verify the correct blinking agent initial in the monitor window.
Is it possible to adjust the left position vapor? ___ (N)
Is it possible to adjust the right position vapor? ___ (N).
Adjust the handwheel on the middle position vapor to zero (0).
Adjust the handwheel on the right position to any concentration
above zero (0).
Verify that the yellow indicator lamp above the right position
vaporizer is illuminated. Press AGENT SELECT on the multispec
monitor and verify the correct blinking agent initial in the monitor
window.
Is it possible to adjust the left position vapor? ___ (N).
Is it possible to adjust the middle position vapor? ___ (N).
Adjust the handwheel on the right position vapor to zero (0).
6-53
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)