Dräger Narkomed 3 Technical & Service Manual page 262

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

PMC PROCEDURE (continued)
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
6.39.5.4
Attach the CO
4110216 A) to the CO
4107979 P)
6.39.5.5
Turn the flow control valve on the calibration adapter
slightly counter-clockwise.
6.39.5.6
After the CO
turn off the flow and remove the sample line from the
calibration adapter.
6.39.5.7
Verify that the ET CO
to 40.0 mm Hg.
6.39.5.8
Press the monitor Display key to scroll to ET AGENT.
6.39.5.9
Verify that the "ET" halothane on the agent display
indicates within 0 to 0.4%
6.39.5.10
Verify that the "EXP" N
38.0% on the data CRT.
6.39.5.11
Press the TRACE keypad. Using the arrow keys below
the Trace CRT, select the HALOTHANE scale.
6.39.5.12
Ensure that all vaporizer handwheels are set to the
zero, or OFF position.
6.39.5.13
Fill each vaporizer with the correct anesthetic agent
until the liquid level is within the correct range as
indicated on the sight glass. DO NOT OVER FILL.
6.39.5.14
Calibrate a Riken gas indicator Model 18 as per
manufacturer's instructions.
6.39.5.15
Insert the patient circuit elbow into the fresh gas
outlet.
6.39.5.16
Connect a sampling "T" between the elbow and the
fresh gas hose.
6.39.5.17
Turn the APL valve knob fully counter-clockwise to its
open position.
6.39.5.18
Attach a short 22 mm hose between the inspiratory
and expiratory valves.
6.39.5.19
Attach a breathing bag to swivel bag mount connector.
6-56
RETURN TO CD-ROM TABLE OF CONTENTS
calibration adapter assembly (P/N
2
calibration cylinder (P/N
2
waveform peaks on the remote display,
2
display indicates within 36.0
2
O indicates within 32.0 to
2
NM3
Rev. V