Dräger Narkomed 3 Technical & Service Manual page 251

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
6.35
Sphygmomed
6.35.1
6.35.2
6.35.3
6.35.4
6.35.5
6.35.6
6.35.7
6.35.8
6.35.9
6.35.10
6.35.11
6.35.12
6.35.13
6.35.14
Rev. V
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Press the START key and verify that the green LED is lighted,and
that the pump begins to inflate the cuff. After the monitor detects
that the cuff is either disconnected, applied wrong, or has a leak,
the pump should stop. The Start LED should extinguish, and the
yellow BP ERROR should be lighted.
The CRT shall display a "BP CUFF ERR" message under the
Advisory column.
Turn the System Power switch to STANDBY. Wrap the cuff loosely
around a cylindrical object. Attach, in series, a calibrated
manometer between the patient cuff and the hose assembly.
Press and hold the
switch to ON.
The cuff shall inflate to 180 to 190 mm Hg in 4 seconds or less. The
manometer shall indicate ±2 mm Hg, and the pressure shall not
drop more than 3 mm Hg before the system deflates. The cuff
should deflate 29 to 31 seconds after the start of inflation. Remove
the cuff from the cylindrical object.
Turn the System Power switch to STANDBY.
Turn the System Power switch to ON.
Install the BP cuff on your left arm.
Press the DISPLAY key and set the interval to 5 minutes.
Press the START key.
When the cuff cycle is complete, press the LO key and set the alarm
limit to above the actual value. Decrease the setting to below actual
value; the alarm should cease.
Press the HI key and set the alarm limit to below actual value.
Increase the setting to above actual value; the alarm should cease.
Press the STAT key. The cuff should inflate. The time between
deflation and the next inflation shall be 3 seconds.
Press the STOP key.
6-47
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
and
keys, and turn the System Power