Dräger Narkomed 3 Technical & Service Manual page 193

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
5.10.5
5.10.6
5.10.7
5.10.8
5.10.9
NOTE: The adjustment range is 600 to 800 mm Hg.
5.10.10 Press the DISPLAY key to enter the value into non-volatile memory.
5.10.11 Cycle the System Power switch to STANDBY and back to ON.
5.10.12 Verify the Capnomed Analyzer voltages: Refer to Field Service
5.10.13 Verify the Capnomed Analyzer flow rate: Refer to Field Service
5.10.14 Close the monitor housing back cover and replace the retaining
5.10.15 Perform the PMS Procedure given in Section 6.
'BLANK' KEY
Figure 5-12: CAPNOMED MONITOR
Rev. B
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
Turn the System Power switch to STANDBY.
Turn the System Power switch to ON while simultaneously pressing
the
and
keys on the Capnomed (B) monitor.
After approximately ten seconds, a "10" will appear in the monitor
display window.
Press the "blank" key to the right of the DISPLAY key (see Figure 5-
12).
A number between 600 and 800 should appear on the monitor
display. Using the
nominal barometric pressure reading obtained in Step 5.10.4
Procedure SP00051, NM3 Capnomed Analyzer Voltage Adjustment.
Procedure SP00052, NM3 Capnomed Analyzer Flow Adjustment.
screws. Re-connect any external cables that were previously removed
from the serial ports.
ET CO
2
INSP CO
2
APNEA
CO
HI
2
CO
LO
2
LINE BLOCK
5-33
RETURN TO CD-ROM TABLE OF CONTENTS
and
keys, adjust the display to show the
DISPLAY
END TIDAL CO
2
HI
LO
CO
2
CO
CAL
ALARMS
2
DISABLE
ALARMS