Dräger Narkomed 3 Technical & Service Manual page 143

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
4.32.5
Remove the plastic cap in the side of the frame rail to provide access
to the caster stem set screw.
4.32.6
Loosen the set screw and remove the caster.
4.32.7
Insert the replacement caster into the frame and hold it in its seated
position.
4.32.8
Tighten the caster stem set screw and replace the plastic cap in the
frame rail.
4.32.9
Using at least two people, tilt the machine, remove the support brace
and carefully lower the machine to the floor.
4.32.10 Check for proper operation of the caster and ensure that the front
casters lock properly.
4.32.11 Replace any unsecured equipment and accessories that were
previously removed.
4.32.12 Perform the PMS Procedure given in Section 6, including a vaporizer
calibration verification.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
REPLACEMENT PROCEDURES (continued)