Dräger Narkomed 3 Technical & Service Manual page 216

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

PMC PROCEDURE (continued)
( ) 6.8 Oxygen Concentration Test
6.8.1
Oxygen + Nitrous Oxide Concentration Test
6.8.1.1
6.8.1.2
6.8.1.3
6.8.1.4
6.8.1.5
6.8.1.6
6.8.1.7
6.8.1.8
6.8.1.9
6.8.1.10 What is the oxygen concentration? ___ % (64-70)
6.8.1.11 Close the oxygen and nitrous oxide flow control valves.
6.8.2
Oxygen + Air Concentration Test - If Applicable
6.8.2.1
6.8.2.2
6.8.2.3
6.8.2.4
6.8.2.5
6.8.2.6
6.8.2.7
6.8.2.8
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Connect a 12-inch hose to the inspiratory valve.
Set the Manual/Automatic to BAG.
Close the APL valve.
Occlude the bag mount.
Insert the sensor from a calibrated O
adapter on the inspiratory valve.
Depress the O
FLUSH button for 15 seconds.
2
Set the oxygen flow to 4 lpm.
Does the O
Med read 97-100% within 3 minutes? ___ (Y)
2
Set the nitrous oxide flow to 2 lpm.
Set the gas selector switch to ALL GASES.
Depress the O
FLUSH button for 15 seconds.
2
Does the O
Med read 97-100% within 3 minutes? ___ (Y)
2
Set the oxygen flow to 4lpm.
Set the air flow to 2 lpm.
What is the oxygen concentration? ___ % (71-77)
Close the oxygen and air flow control valves.
Set the gas selector switch to O
6-14
RETURN TO CD-ROM TABLE OF CONTENTS
Med into the valve dome
2
+N
O.
2
2
NM3
Rev. V