Dräger Narkomed 3 Technical & Service Manual page 267

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
6.45
Oxygen Flush Test
6.45.1
6.45.2
6.45.3
6.45.4
6.45.5
( )
6.45.6
6.45.7
6.45.8
6.45.9
6.45.10
( )
6.45.11
6.45.12
6.45.13
6.45.14
Rev. V
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Turn the SYSTEM POWER switch to STANDBY.
Press and release the O
Does the flow of oxygen stop immediately? __ (Y)
Using a 15 mm male connector and short piece of tubing, connect a
test minute volumeter to the common gas outlet.
Press the O
FLUSH button.
2
What is the oxygen flush flow rate? ___ lpm. (45-65)
Remove the test minute volumeter.
Turn the SYSTEM POWER switch to ON.
Insert the sensor from a calibrated O
dome.
Press the O
FLUSH button.
2
What is the O
concentration? ___ % O
2
Remove and cap the O
adapter.
Close the oxygen cylinder valve.
Bleed the oxygen circuit by pressing the O
6-61
RETURN TO CD-ROM TABLE OF CONTENTS
PMC PROCEDURE (continued)
FLUSH button.
2
Med into the inspiratory valve
2
(97-100)
2
Med sensor; plug the inspiratory valve dome
2
FLUSH button.
2