Dräger Narkomed 3 Technical & Service Manual page 27

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

NM3
OTHER ERROR
CODES IN MONITOR
WINDOW?
Y
REPLACE MONITOR
AS OUTLINED IN
PROCEDURE 4.22
DOES MONITOR
OPERATE PROPERLY?
N
REINSTALL
ORIGINAL MONITOR
CONTACT NAD
SERVICE DEPARTMENT
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
TROUBLESHOOTING GUIDE (continued)
GUIDE 7 (continued): MULTISPEC Monitor and Analyzer Errors
LOW END
N
TIDAL VOL AND INSP
CO2 READINGS?
Y
RECALIBRATE
MULTISPEC ANALYZER
LOW END
TIDAL VOL AND INSP
Y
CO2 READINGS?
Y
REPLACE
MULTISPEC ANALYZER
AS OUTLINED IN
PROCEDURE 4.19
DOES ANALYZER
OPERATE PROPERLY?
N
REINSTALL ORIGINAL
MULTISPEC ANALYZER
3-13
RETURN TO CD-ROM TABLE OF CONTENTS
N
SLOW
RESPONSE TIME?
Y
CHECK FOR
SAMPLE LINE
OCCLUSION
SLOW
N
RESPONSE TIME?
Y
PERFORM A
Y
COMPLETE PMS
ON THE MACHINE
DOES
N
MACHINE PASS
PMS?
MACHINE IS
FUNCTIONAL
N
N
Y
DOES ANALYZER
OPERATE PROPERLY?
N
Y