PMC PROCEDURE (continued)
6.9.5
Carbon Dioxide Flowmeter Test - If Applicable
6.9.5.1
Set the gas selector switch to ALL GASES.
6.9.5.2
Open the carbon dioxide gas cylinder valve.
6.9.5.3
Is it possible to adjust the flow of the carbon dioxide over the full
range of the flowmeter? ___ (Y)
6.9.5.4
Close the carbon dioxide flow control valve.
6.9.5.5
Is the correct flow control knob and label attached to the carbon
dioxide flow control valve? ___ (Y)
6.9.5.6
Set the gas selector switch to O
6.9.6
Auxiliary Oxygen Flowmeter Test - If Applicable
6.9.6.1
Connect a test cm H
6.9.6.2
Bleed any pressure.
6.9.6.3
Is there an increase in pressure? ___ (N)
6.9.6.4
Increase the pressure to 50 cm H
6.9.6.5
After thirty (30) seconds, what is the pressure? ___ cm H
6.9.6.6
Remove the test gauge.
6.9.6.7
Is it possible to adjust the flow over the full range of the flowmeter?
___ (Y)
6.9.6.8
Adjust the flow to 5 lpm.
6.9.6.9
Hold the sensor from a calibrated O
6.9.6.10 What is the oxygen concentration? ___ % (97-100)
6.9.6.11 Remove the O
6.9.6.12 Close the flowmeter flow control valve.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
+N
2
2
O gauge to the outlet.
2
O.
2
Med sensor.
2
6-18
O.
2
Med at the flowmeter outlet.
2
NM3
O (30-50)
Rev. V