Dräger Narkomed 3 Technical & Service Manual page 268

Anesthesia system
Hide thumbs Also See for Narkomed 3:
Table of Contents

Advertisement

PMC PROCEDURE (continued)
( )
6.46
Operator's Manual
Verify the availability/location of the NARKOMED 3 Operator's Manual.
( )
6.47
Final Check
6.47.1
6.47.2
6.47.3
6.47.4
6.47.5
6.47.6
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
Verify that the pipeline hoses are connected to the hospital pipeline.
Verify
that
the
APL
counterclockwise, fully open.
Verify that the O
Med sensor is removed from the valve dome
2
adapter.
Verify that the valve dome is plugged.
Verify that the machine is plugged into a live outlet.
Verify that the Manual/Automatic selector valve is in the MANUAL
position.
6-62
RETURN TO CD-ROM TABLE OF CONTENTS
valve
knob
is
turned
NM3
completely
Rev. V