Anritsu Spectrum Master MS2722C Maintenance Manual page 94

High performance handheld spectrum analyzer
Hide thumbs Also See for Spectrum Master MS2722C:
Table of Contents

Advertisement

Chapter 3 — Options Verification
11. Adjust the MG3700A Level setting to approximately –15 dBm so that the power meter reads
–30.0 dBm ± 0.5 dB.
12. Record the MS272xC Residual EVM(rms) and Frequency Error reading into the section
"750 MHz, –30 dBm" in
13. Verify that the values are within specification.
14. Set the calibration factor frequency of the power sensor to 2350 MHz.
15. Set MG3700A Frequency to 2350 MHz. Press the Set key.
16. Adjust the MG3700A Level setting with the knob so that the power meter reads –12.0 dBm ± 0.5 dB.
17. Change the Center Frequency of the MS272xC to 2350 MHz.
18. Record the MS272xC Residual EVM and Frequency Error reading into the section
"2350 MHz, –12 dBm" in
19. Verify that the values are within specification.
20. Adjust the MG3700A Level setting to approximately –18 dBm so that the power meter reads
–30.0 dBm ± 0.5 dB.
21. Record the MS272xC Residual EVM and Frequency Error reading into the section
"2350 MHz, –30 dBm" in
22. Verify that the values are within specification.
23. For instruments with Option 543, press the Setup main menu key and set the BW to 20 MHz. Then
repeat
Step 5
Substep
Step 5
Substep i.
MS272xC MM
3-11 TD-LTE Signal Analyzer Option 551 and/or 552 and/or 543
Table
A-129.
Table
A-129.
Table
A-129.
a
through
Step 22
by using the TDLTE-E-TM-3-3_20M pattern for
PN: 10580-00279 Rev. E
3-49

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Spectrum Master MS2722C and is the answer not in the manual?

Table of Contents