A-2 Test Records for Spectrum Analyzer Verification
MS272__C
Firmware Rev: ______________ Operator: ____________________
Serial Number: _______________
The listed Amplitude Accuracy specifications apply when the instrument is being verified in an
environment between 20°C and 30°C after a 30 minute warm-up period. For an instrument that is not
being verified in a temperature controlled environment, but within a temperature range between
Note
–10°C and 50°C (–10°C and 40°C for MS2726C only) and after a 60 minute warm-up, add ± 1 dB to
the Amplitude Accuracy specification for frequencies from 100 kHz to 32 GHz, add ± 2 dB to the
Amplitude Accuracy specification for frequencies from 32 GHz to 40 GHz.
Table A-56. Verification at 38.0 GHz, Specification = ± 2.3 dB
Input Power
–30 dBm
–30 dBm
–30 dBm
–30 dBm
–5 dBm
–5 dBm
–5 dBm
–5 dBm
Table A-57. Verification at 39.0 GHz, Specification = ± 2.3 dB
Input Power
–30 dBm
–30 dBm
–30 dBm
–30 dBm
–5 dBm
–5 dBm
–5 dBm
–5 dBm
A-30
Options: __________________________________________________
Atten Lvl
0 dB
5 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
Atten Lvl
0 dB
5 dB
10 dB
20 dB
30 dB
40 dB
50 dB
60 dB
PN: 10580-00279 Rev. E
Appendix A — Test Records
Marker 1 Reading
Marker 1 Reading
Date: ____________
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
dBm
MS272xC MM