Channel Power Accuracy (10 Mhz Bandwidth And 10 Ms Frame Length) - Anritsu Spectrum Master MS2722C Maintenance Manual

High performance handheld spectrum analyzer
Hide thumbs Also See for Spectrum Master MS2722C:
Table of Contents

Advertisement

Chapter 3 — Options Verification
Setup
1. Connect the Power Sensor to the power meter and zero the sensor.
2. Set the Power Meter Measurement MODE to True RMS, set Averaging MODE to Moving, and set
Averaging NUMBER to 256.
3. Set the calibration factor frequency of the power sensor to 2600.5 MHz.
4. Connect all of the test equipment as shown in

Channel Power Accuracy (10 MHz Bandwidth and 10 ms Frame Length)

Procedure
1. Set the MG3700A as follows:
a. Press the yellow Preset button (answer yes to the question).
b. Press the Set key
c. Press the (F1) submenu key to select "Load File to Memory".
d. Press the (F1) submenu key again to select "Select Package".
e. Using the Down Arrow key, step through the selection list until mWiMAX is highlighted.
f. Press the Set key.
g. Press the (F6) submenu key, Return.
h. Press the Set key. The "Select Package" list box is displayed. Again select mWiMAX and press Set.
i. Another file list is displayed. Select (highlight) "10m1024g8_0_10_cap".
j. Press the Set key.
k. Press the MOD On/Off key and verify that the LED is On. Ensure that the "Playing" indicator is
displaying the moving pattern.
l. Press the Frequency key, then enter 2600.5 MHz.
m. Press the Level key, then enter 2 dBm. Turn the output On.
2. Adjust the MG3700A Level setting with the knob so that the power meter reads – 15.0 dBm ± 0.2 dB, and
record the actual value as shown on the power meter into
Channel Power Accuracy Power Error (2600.5 MHz and –15.0 dBm)" on page A-50.
3. Set the MS272xC to Mobile WiMAX Signal Analyzer mode and Preset the instrument.
4. Set the MS272xC as follows:
a. Press the Freq main menu key and set the Center Freq to 2600.5 MHz.
b. Press the Setup main menu key and set the Bandwidth to 10 MHz.
c. Press the Frame Length submenu key and set the Frame Length to 10 ms.
d. Press the Measurements main menu key, press RF, and then press Power vs. Time.
5. Record the "MS272xC Reading", the displayed Channel Power (RSSI) value, into
6. Calculate the Channel Power Error by subtracting the "MS272xC reading" from the "Power Meter
Reading" in
Step
7. Verify that the error is within specification.
8. Adjust the MG3700A Level setting to approximately –33 dBm so that the power meter reads
–50.0 dBm ± 0.2 dB, and record the actual value as shown on the power meter into
"Option 66, Mobile WiMAX Channel Power Accuracy Power Error (2600.5 MHz and –50.0 dBm)"
on page A-50.
MS272xC MM
3-9 Option 66 and/or Option 67, Mobile WiMAX Signal Analyzer
2. Record the result into
PN: 10580-00279 Rev. E
Figure
3-10.
Table A-114, "Option 66, Mobile WiMAX
Table
A-114.
Table
A-114.
Table A-115,
3-35

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Spectrum Master MS2722C and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents