Demo #2 - Magnitude Cal With Source Control, Multiple States - Keysight Technologies U9361C Measurement Manual

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Demonstrations
Demo #2 – Magnitude Cal with Source Control, Multiple States
Compatible Signal Generators:
— N5173B EXG
— N5183B MXG
— N519xA (all UXG)
This demonstration shows how changing measurement parameters after a calibration may require
adding additional Cal states. A signal generator and a signal analyzer connected through USB is
required for source control. This demo calibrates for a 3 dB attenuator. For signal generator control,
refer to
"Setting Up a Signal Analyzer to Control a Signal Generator" on page
21.
On the Signal Analyzer:
1. Select Mode > Spectrum Analyzer > Mode Preset to set the signal analyzer to a known state.
2. Select FREQ > Start Freq > 24 GHz
3. Stop Freq > 25 GHz
28
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide