U3047AM08
Control Lines

Setting the Control Lines with Address and Data Values

This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the network analyzer
is in its Standalone Mode.

Setting the Network Analyzer to Standalone Mode

1. Select Utility > System > Configure > Multiport Capability -OR-
As described in
page
with the following exceptions:
— The <address> value is always 112
Test Set DUT Control Address and Data Logic Table
Table 13
<Address>
<Data>
112
112
112
112
112
112
112
112
112
112
Keysight U3047AM08 User's and Service Guide
(Instrument > Setup > External Hardware > Multiport > Multiport
Configuration...). In the dialog box, select Restart as a standalone PNA > OK.
"Setting the Test Port Paths with Address and Data" on
47, the <address>.<data> values are determined in a similar manner,
Table 13
will be used
Description
0
ALL DUT Control Lines set to logic high or connected to Pin 13
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1
DUT Control Line 1 set to logic low or connected to Pin 14
2
DUT Control Line 2 set to logic low or connected to Pin 14
4
DUT Control Line 3 set to logic low or connected to Pin 14
8
DUT Control Line 4 set to logic low or connected to Pin 14
16
DUT Control Line 5 set to logic low or connected to Pin 14
32
DUT Control Line 6 set to logic low or connected to Pin 14
64
DUT Control Line 7 set to logic low or connected to Pin 14
128
DUT Control Line 8 set to logic low or connected to Pin 14
Line/Pin
1-8
1-8
1
2
3
4
5
6
7
8
53