13.Select Input Output > Calibration > Configuration to navigate to the RCal Calibration
Configuration window.
14.Select Use Current Meas > Calibrate > Calibrate Checked Rows to populate a row using the
current measurement settings and run the calibration.
15. When the calibration is complete, select the Apply box.
16. Disconnect the cable from the RCal module and connect it to the RF output of the signal
generator.
17. Select Source > RF Output > On to enable the signal generator output.
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
Demonstrations
31