Setting the Control Lines with Address and Data Values

This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in
its Standalone mode.

Setting the Analyzer to Standalone Mode

1. Select Utility > System > Configure > Multiport Capability -OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...).
In the dialog box select Restart as a standalone PNA > OK.
2. Access Interface Control Mode (see
As described in
"Setting the Individual Test Port Paths with Address and Data" on page
<address>.<data> values are determined in a similar manner, with the following exceptions:
• Table 10 will be used.
• The <address> value is always = 112
Table 11
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
112
0
112
255
112
1
112
2
112
4
112
8
112
16
112
32
112
64
112
128
54
page
39)
Description
ALL DUT Control Lines set to logic high or connected to Pin 13
ALL DUT Control Lines set to logic low or connected to Pin 14
DUT Control Line 1 set to logic low or connected to Pin 14
DUT Control Line 2 set to logic low or connected to Pin 14
DUT Control Line 3 set to logic low or connected to Pin 14
DUT Control Line 4 set to logic low or connected to Pin 14
DUT Control Line 5 set to logic low or connected to Pin 14
DUT Control Line 6 set to logic low or connected to Pin 14
DUT Control Line 7 set to logic low or connected to Pin 14
DUT Control Line 8 set to logic low or connected to Pin 14
Keysight U3022AM42 User's and Service Guide
Control Lines
45, the
Line/Pin
1-8
1-8
1
2
3
4
5
6
7
8