Keysight Technologies U9361C Measurement Manual page 42

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Demonstrations
19.Calibrate and apply the new state as previously demonstrated for the uW path.
20.Disconnect the cable from the RCal module and connect it to the signal generator. Enable the
RF output on the signal generator.
RF Output > On
The EVM is now reduced to 0.12% with the Standard path. By leaving both paths applied, the
instrument will automatically select the proper calibration state.
42
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide