U3042AM08
Control Lines
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
NOTE
The following control feature will function only while the analyzer is in Standalone
Mode.
Setting the Analyzer to Standalone Mode
1. Select Utility > System > Configure > Multiport Capability -OR-
As described in
are determined in a similar manner, with the following exceptions:
•
•
Table 16
Test Set DUT Control Address and Data Logic Table
<Address>
112
112
112
112
112
112
112
112
112
112
62
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...).
In the dialog box, select Restart as a Standalone PNA > OK
"Address and Data Values" on page
Table 16
will be used.
The <address> value is always = 112
<Data>
Description
0
ALL DUT Control Lines set to logic high or connected to Pin 13
255
ALL DUT Control Lines set to logic high or connected to Pin 14
1
DUT Control Line 1 set to logic low or connected to Pin 14
2
DUT Control Line 2 set to logic low or connected to Pin 14
4
DUT Control Line 3 set to logic low or connected to Pin 14
8
DUT Control Line 4 set to logic low or connected to Pin 14
16
DUT Control Line 5 set to logic low or connected to Pin 14
32
DUT Control Line 6 set to logic low or connected to Pin 14
64
DUT Control Line 7 set to logic low or connected to Pin 14
128
DUT Control Line 8 set to logic low or connected to Pin 14
56, the <address>.<data> values
Keysight U3042AM08 User's and Service Guide
Line/Pin
1-8
1-8
1
2
3
4
5
6
7
8