Keysight Technologies U9361C Measurement Manual page 53

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Frequency (GHz)
Power Sensor (dBm)
17
-10.12
20
-10.36
The average difference in results between the Power Sensor and the measurement using an RCal
calibration for this test case was 0.192 dBm across the selected measurement span.
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
RCal Corrected
Difference (Abs)
-10.08
0.04
-10.01
0.35