Demonstrations
Calibrate > Calibrate Checked Rows
25.When calibration is complete, click the Apply box.
26.Disconnect the cable from the RCal module and connect it to the signal generator. Enable the
RF output on the signal generator through Meas Setup > Source >RF Output > On.
Now the trace should be properly corrected and without errors.
27.View the new trace by adding a 3rd trace.
Trace > Select Trace > Trace 3
Trace Type > Max Hold
View > Active
34
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide