Return > Symbol Rate > 10 Msps
Return > Digital Modulation ON
RF > ON
4. Open the Digital Demod application in Vector Modulation Analyzer mode.
Mode > Vector Modulation Analyzer > Digital Demod
5. Configure the signal analyzer to demodulate the waveform.
Ctr (Center Frequency) > 5GHz
Format > 128-QAM
Sym Rate > 10MHz
Info BW > 20MHz
You should now see the constellation, signal, and metrics.
6. On the signal generator, enable trace averaging to smooth the signal.
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
Demonstrations
37