Keysight Technologies U9361C Measurement Manual page 38

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Demonstrations
Meas Setup > Averaging > On
7. Disable the RF output on the signal generator.
RF Output > Off
8. Disconnect the cable from the signal generator and connect it to the RCal module.
9. Navigate to the RCal Calibration Configuration window.
Input/Output > Calibration > Configuration
The Calibration Configuration table should be unpopulated.
38
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide