U3042AM04
Control Lines

Setting the Control Lines with Address and Data Values

This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in Standalone
Mode.

Setting the Analyzer to Standalone Mode

1. Select Utility > System > Configure > Multiport Capability -OR-
As described in
are determined in a similar manner, with the following exceptions:
Table 14
Test Set DUT Control Address and Data Logic Table
<Address>
112
112
112
112
112
112
112
112
112
112
Keysight U3042AM04 User's and Service Guide
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...).
In the dialog box select Restart as a standalone PNA > OK.
"Address and Data Values" on page
Table 14
will be used.
The <address> value is always = 112
<Data>
Description
0
ALL DUT Control Lines set to logic high or connected to Pin 13
255
ALL DUT Control Lines set to logic high or connected to Pin 14
1
DUT Control Line 1 set to logic low or connected to Pin 14
2
DUT Control Line 2 set to logic low or connected to Pin 14
4
DUT Control Line 3 set to logic low or connected to Pin 14
8
DUT Control Line 4 set to logic low or connected to Pin 14
16
DUT Control Line 5 set to logic low or connected to Pin 14
32
DUT Control Line 6 set to logic low or connected to Pin 14
64
DUT Control Line 7 set to logic low or connected to Pin 14
128
DUT Control Line 8 set to logic low or connected to Pin 14
40, the <address>.<data> values
Line/Pin
1-8
1-8
1
2
3
4
5
6
7
8
46