Control Lines

Setting the Control Lines with Address and Data Values

This section describes how to control the DUT control lines. Refer to
Lines" on page

Setting the PNA to Standalone Mode

1. Select Utility > System > Configure > Multiport Capability -OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...).
In the dialog box, select Restart as a Standalone PNA > OK.
As described in
<address>.<data> values are determined in a similar manor, with the following
exceptions:
Table 15
— The <address> value is always = 112
All DUT control lines remain in a high logic state until set by the address data 0 = logic
high. All control lines are set by the sum of the data sent to the Test Set.
Example: To set control line 1 and 8 to a low state send command 112.129 (1 + 128).
All other lines remain in a high state.
Table 15
Address
112
112
112
112
112
112
112
112
112
112
58
54.
"Setting the Test Port Paths with Address and Data" on page
will be used.
Test Set DUT Control Address and Data Logic Table
Data
Description
0
ALL the Control Lines set to logic high
255
ALL the Control Lines set to logic low
1
DUT Control Line 1 set to logic low
2
DUT Control Line 2 set to logic low
4
DUT Control Line 3 set to logic low
8
DUT Control Line 4 set to logic low
16
DUT Control Line 5 set to logic lowe
32
DUT Control Line 6 set to logic low
64
DUT Control Line 7 set to logic low
128
DUT Control Line 8 set to logic low
Keysight U3042AM20 User's and Service Guide