Key Steps for Every RCal Calibration

The following procedure should be followed before every RCal calibration to guarantee the best
possible accuracy:
1. RCal calibrations are independent of measurement time and temperature, which are
parameters that are handled through routine spectrum analyzer alignments. For the best
possible accuracy, alignments should be run before every RCal calibration and after a 30
minute warm-up time. The RCal module does not replace spectrum analyzer calibrations and
alignments.
2. Make sure that the Ext 10 MHz reference is connected and that the selected USB connection
supplies ample current.
3. If source control is desired for the calibration, connect the necessary input and output triggers
between the analyzer and source.
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
RCal Software – Calibration and Application
19