12.Populate a row using the current measurement settings. Duplicate this row if comparisons for
both the Standard Path and
Use Current Meas > Duplicate Row > µW Path Control > µW Presel Bypass
13.Run the calibration after selecting the calibrate parameter.
Calibrate > Calibrate > Checked Rows
14.When calibration is complete, select Apply.
15.Disconnect the cable from the RCal module and connect it to the signal generator. Enable the
RF output on the signal generator.
RF Output > On
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
µ
W Preselector Bypass are desired.