Demonstrations
Demo #1 – Simple Magnitude Calibration
Compatible Signal Generators:
— N5166B (CXG)
— N517xB (all EXG)
— N518xB (all MXG)
— N519xA (all UXG)
— M9x84B/C (all VXG)
This demonstration corrects for loss introduced by a 13 dB attenuator in the signal path. Before
beginning this demo, please follow the guidelines detailed in the
Calibration" on page
19.
1. Connect the RF output of the signal generator to the RF input of the signal analyzer with the
attenuator in between.
On the Spectrum Analyzer:
2. Select Mode Preset to set the analyzer to a known state.