Keysight Technologies U9361C Measurement Manual page 24

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Demonstrations
On the signal generator: Configure the following settings to sweep across the defined
measurement span.
1. Select Preset to set the signal generator to a known state.
2. Set Amptd > -10 dBm.
3. Select Sweep > Sweep > Freq > On to enable frequency sweeping.
4. Select Return> Configure Step Sweep > Freq Start > 3 GHz and Freq Stop > 4 GHz.
5. Set # Points > 101.
6. Select More > Step Dwell > 2.00 msec.
7. Set RF Out > On. The RF output state is indicated by the LED next to the toggle button.
On the signal analyzer:
1. Select Trace. In the Select Trace drop-down, select Trace 1.
2. Set Trace Type to Max Hold, and then in the View/Blank area, select Active.
24
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide