Keysight Technologies U9361C Measurement Manual page 40

Rcal receiver calibrator
Hide thumbs Also See for U9361C:
Table of Contents

Advertisement

Demonstrations
Calibrate > Calibrate Checked Rows
12.When calibration is complete, click the Apply box.
13.Disconnect the cable from the RCal module and connect it to the signal generator. Enable the
RF output on the signal generator.
RF Output > On
The EVM should now be improved, as shown in the comparison below.
14.Change the RF path to use the Yig Tuned Filter (Standard Path).
µW Path > Standard
This will generate a warning, since the current corrections only apply to the
40
U9361C/F/G/M RCal Receiver Calibrator Measurement Guide
µ
W Bypass Path.