U3022AM06
Control Lines

Setting the Control Lines with Address and Data Values

This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in
Standalone Mode.

Setting the Network Analyzer to Standalone Mode

1. Select Utility > System > Configure > Multiport Capability -OR-
As described in
<address>.<data> values are determined in a similar manor, with the following
exceptions:
Table 18 Test Set DUT Control Address and Data Logic Table
<Address>
112
112
112
112
112
112
112
112
112
112
68
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...).
In the dialog box, select Restart as a Standalobe PNA > OK.
"Setting the Test Port Paths with Address and Data" on page
Table 18
will be used.
The <address> value is always = 112
<Data>
Description
0
ALL DUT Control Lines set to logic high or connected to Pin 13
255
ALL DUT Control Lines set to logic high or connected to Pin 14
1
DUT Control Line 1 set to logic low or connected to Pin 14
2
DUT Control Line 2 set to logic low or connected to Pin 14
4
DUT Control Line 3 set to logic low or connected to Pin 14
8
DUT Control Line 4 set to logic low or connected to Pin 14
16
DUT Control Line 5 set to logic low or connected to Pin 14
32
DUT Control Line 6 set to logic low or connected to Pin 14
64
DUT Control Line 7 set to logic low or connected to Pin 14
128
DUT Control Line 8 set to logic low or connected to Pin 14
Keysight U3022AM06 User's and Service Guide
60, the
Line/Pin
1-8
1-8
1
2
3
4
5
6
7
8