7.2 Fault Chart
FAULT
FAULT 01 – Not used.
Alarm sounds, and display shows:
FAULT 02
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 02
FAULT 03 – Not used.
*Service charges may be incurred. Consult your warranty for details.
Troubleshooting
IMPORTANT: When a situation occurs with the Reliance EPS
Processor, an alarm will sound. Refer to S
F
TO A
IMPORTANT: When a fault condition or situation occurs, the cycle
automatically stops. Endoscopes may have to be reprocessed
through a complete Reliance EPS Cycle.
1. Building or processor hot water supply valve is not fully
2. Building supply-line strainers are clogged. Verify strainer;
3. Dual Pre-Filter and Pressure Regulator Assembly (required
MM-DD-YY
=
4. Air bubble in hot water supply lines.
5. Building hot water supply pressure is too low (refer to
6. Other possible causes:
7. Call STERIS.*
Operator Manual
, on how to respond to faults.
AULT
POSSIBLE CAUSE AND CORRECTION
SUMP HTW FILL TOO LONG
NOTE: HTW = Hot Tap Water.
open.
•
Open hot water supply valves. Press
pad to resume cycle.
clean if necessary.
accessory) needs service.
a. Verify last ten cycle printouts. An increase of maximum
filling time indicates Dual Pre-Filter and Pressure
Regulator Assembly cartridges are clogged. Change
filter cartridges (refer to S
b. Perform a cycle. If a situation reoccurs, change
0.2 μm bacterial-retentive filter cartridge (refer to
S
6.10.2).
ECTION
c. If situation reoccurs, see Step 4.
a. Press START touch pad to resume cycle. Processor
will be drained and should evacuate any air present.
b. If situation reoccurs, see Step 5.
equipment drawing [920-011-656]). Call STERIS.*
-
Defective hot water fill valves.
-
Defective water level switch.
-
Defective drain valve.
4.13, R
ECTION
ESPONDING
START
6.10.1).
ECTION
920011-649
touch
7-5