Cycle and Control Value Programming
19. Press ENTER touch pad to confirm setting. Display shows:
20. Physician ID function allows operator to scan a bar code to
identify physician who used endoscope to be processed.
Physician ID is recorded on cycle printout.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) optional Physician ID function.
21. Press ENTER touch pad to confirm setting. Display shows:
22. This screen is to set day countdown before next change of
0.2 μm bacterial-retentive filter. Default setting is 90 days. Range
is between 0 and 90 days.
This setting may be changed depending on use frequency of
processor and quality of water. It also must be reset every time
filter is replaced. After a few months of observing alarm pattern,
supervisor may decide to shorten delay as a preventive measure
to be able to change filter before FAULT 30 (0.2 μm FILTER
FAILURE) alarm is triggered, or to prevent filling time from
becoming too long (see S
details on fault causes and solutions).
To change setting, press SCROLL arrows (left or right) to scroll
to numbers and set day countdown.
23. Press ENTER touch pad to confirm setting. Display shows:
24. WASHING SELECT screen allows operator to toggle WITH
WASHING/WITHOUT
endoscope processing cycle or a DOUBLE endoscope
processing cycle is selected.
Press SCROLL arrows (up or down) to enable or disable
WASHING SELECT feature (YES = Operator can elect whether
or not to process a WASHING phase; NO = Operator cannot
elect whether or not to process a WASHING phase).
Operator Manual
PHYSICIAN ID
YES
__ blinking
0.2 ΜM FILTER
IN XX DAYS
__ blinking
7, T
ECTION
ROUBLESHOOTING
WASHING SELECT
YES
__ blinking
WASHING
mode
, for
when
a
SINGLE
5-11