Steris RAS 12 Rack Manual
Hide thumbs Also See for RAS 12 Rack:

Advertisement

Quick Links

©
2022, STERIS Corporation. All rights reserved.
Introduction
FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA
This manual contains important information on the proper use and
routine maintenance of the RAS 12 Rack and the RAS 12 Long Rack.
All personnel involved in the use and routine maintenance of this
equipment must carefully review and comply with the Safety
Precautions and instructions contained in this manual. These
instructions are important to protect the health and safety of personnel
using the RAS 12 Rack to process items in the AMSCO
Single-Chamber Washer/Disinfector; or the RAS 12 Long Rack to
process items in the AMSCO
Disinfector, and should be retained in a conveniently accessible area
for quick reference.
NOTE: Please note the abbreviation "RAS" stands for Robotic
Assisted Surgery.
NOTE: Please note that "RAS Rack" used throughout the manual stands
for both Racks (the RAS 12 Rack as well as the RAS 12 Long Rack).
Advisory
IMPORTANT:
when using and servicing this equipment can be found in S
of this manual. Do not use the equipment until you have become
familiar with this information.
STERIS does not intend, recommend or represent in any way that the
RAS Rack be used in the washer/disinfector for the terminal
disinfection or sterilization of any
Rack is intended only to clean and disinfect reusable
instruments in the washer/disinfector.
When
da Vinci
tissues, such instruments must be terminally processed in
accordance with Good Hospital Practices before each use in human
patients, and also in accordance with all directions specified in the
latest Intuitive Surgical
replaced by use of RAS Rack and robotic cycle as described
herein) as well as in any instrument specific instructions and
supplements that may come with individual instruments.
1
da Vinci
2
Intuitive Surgical is a registered trademark of Intuitive Surgical, Inc.
Operator Manual
A WORD FROM STERIS
®
7053HP Single-Chamber Washer/
A listing of the Safety Precautions to be observed
instruments will be contacting blood or compromised
®2
Reprocessing Instructions (except steps
is a registered trademark of Intuitive Surgical, Inc.
®
ECTION
®1
da Vinci
instruments. The RAS
da Vinci
Printed in Canada.
7052HP
1
i
11748808

Advertisement

loading
Need help?

Need help?

Do you have a question about the RAS 12 Rack and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Steris RAS 12 Rack

  • Page 1 NOTE: Please note the abbreviation "RAS" stands for Robotic Assisted Surgery. NOTE: Please note that "RAS Rack" used throughout the manual stands for both Racks (the RAS 12 Rack as well as the RAS 12 Long Rack). Advisory IMPORTANT: A listing of the Safety Precautions to be observed...
  • Page 2: Indications For Use

    STERIS maintains a global staff of well- equipped, qualified service technicians to provide this service, as well as professional repair services. Please contact STERIS for details.
  • Page 3 Manufactured by: The base language of this STERIS Canada ULC document is ENGLISH. Any 490 boulevard Armand-Paris translation must be made from Québec (Québec) the base language document. Canada G1C 8A3 Sales and Service -- Americas: STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 440-354-2600 •...
  • Page 4 11748808 Operator Manual Introduction FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 5: Table Of Contents

    TABLE OF CONTENTS Section Number Description Page A WORD FROM STERIS......................i Advisory................................i Indications For Use............................ii Trademarks..............................ii Service Information............................ii Certification...............................ii SAFETY PRECAUTIONS ....................1-1 LOADING INSTRUCTIONS ....................2-1 RAS Rack – Main Components......................2-1 da Vinci Instruments ........................... 2-2 Before Processing Instruments......................
  • Page 6 TABLE OF CONTENTS (Cont’d) Section Number Description Page 11748808 Operator Manual Table of Contents FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 7: List Of Illustrations

    LIST OF ILLUSTRATIONS Figure Description Page Section 2: Loading Instructions Figure 2-1. RAS 12 Rack Shown – Components ..................2-1 Figure 2-2. Examples of da Vinci Instruments ..................2-2 Figure 2-3. Connector ..........................2-11 Figure 2-4. Removing Connector......................2-11 Figure 2-5.
  • Page 8 Figure 5-16. Installing Holding Cords and O-Rings................. 5-13 Figure 5-17. RAS 12 Rack (FD219) and RAS 12 Long Rack (FD220) – Part 1 of 2......... 5-16 Figure 5-18. RAS 12 Rack (FD219) and RAS 12 Long Rack (FD220) – Part 2 of 2......... 5-18...
  • Page 9: Safety Precautions

    Regularly scheduled preventive maintenance is required for safe and reliable operation of the washer/ disinfector and this equipment. Contact STERIS to schedule preventive maintenance. Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service personnel. Repairs and adjustments performed by unqualified personnel or installation of unauthorized parts could cause personal injury, result in improper equipment performance, void the warranty or result in costly damage.
  • Page 10 Processing the instruments in the RAS Rack and custom robotic cycle does not replace OR steps, Decon steps or Prepare for Automated Cleaning steps. STERIS makes no claim of effective processing when these instructions are not followed, or are applied to devices that do not match the descriptions and depictions provided.
  • Page 11: Loading Instructions

    2 Additional Ones] Flexible Hoses For Port Connections (X 12) Cool-Touch Handle (X 4) Holding Cords (X 12) Figure 2-1. RAS 12 Rack – Components Shown Loading Instructions Operator Manual 11748808 FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 12: Da Vinci Instruments

    2.2 da Vinci See Figure 2-2. Instruments The RAS Rack can process the following da Vinci instruments: Models da Vinci S/Si Secondary Secondary Flush Port Flush Port Primary Primary Flush Port Flush Port Example of a 5 mm Instrument Example of a 8 mm Instrument Model da Vinci X/Xi Secondary Flush Port...
  • Page 13: Before Processing Instruments

    2.3 Before Processing Before processing da Vinci instruments using the RAS 12 Rack Instruments within the AMSCO ® 7052HP Single-Chamber Washer/Disinfector or the RAS 12 Long Rack within the AMSCO 7053HP Single-Chamber ® Washer/Disinfector, complete the following checklist to help ensure correct equipment operation: ❑...
  • Page 14: Pump Speed

    2. Press Cycles on touch screen. Display shows Supervisor screen: 3. Press Modify Cycles on touch screen. Display shows Cycle Selection screen: 4. Scroll to select any available cycle to modify. Display shows Cycle Modification screen: Proceed with specified settings. NOTE: Any cycle set to locked is not available in cycle start selection.
  • Page 15 5. Press PARAMETERS on touch screen to modify cycle phases. Set the following parameters as indicated: PRE-WASH Phases When settings are completed, press to return to Phase screen Proceed with Wash phase settings... Loading Instructions Operator Manual 11748808 FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 16 WASH Phase NOTE: Ensure to NOTE: CW+RW is assign not allowed to be Enzymatic Cleaner used for this cycle. only for Wash 1 Stage 1. When settings are completed, press to return to Phase screen NOTE: Ensure to assign Neutral Detergent only for Wash 1 Stage 2.
  • Page 17 RINSE Phase When settings are completed, press to return to Phase screen NOTE: PW can alternatively be used for Rinse 1, but it is not allowed if washer is equipped with optional Acu-Rinse Reservoir since phase time cannot be set to desired time in this configuration.
  • Page 18 When settings are completed, press until exiting Supervisor mode. Thermal RINSE Phase Drying Phase IMPORTANT: Drying phase of this cycle is only intended for chamber venting to moisture removal on the exterior of the instruments; interior NOTE: PW can must dried following cycle...
  • Page 19: 2.5 Running Test Cycle

    2.5 Running Test Cycle When all settings are completed, proceed with a Test cycle as follows: 1. Place empty RAS 12 Rack inside chamber. 2. Start robotic cycle. 3. Verify cycle progress and cycle parameters displayed on Cycle Details screen for each phase to confirm proper programming.
  • Page 20: 2.7 Loading Ras Rack

    CAUTION – POSSIBLE EQUIPMENT DAMAGE: To avoid instrument or equipment damage, STERIS does not recommend processing flexible endoscopes, single-use instruments, rigid telescopes (optical lens), fiber-optic cables, multipart instruments that cannot be disassembled, heat-sensitive devices and coagulating electrodes using the RAS Rack in the washer/disinfector.
  • Page 21: Da Vinci Instruments - Model Da Vinci S/Si

    2.7.2 da Vinci Instruments — See Figures 2-3 to 2-11. Model da Vinci S/Si 1. Locate port connectors on instrument holder. Each connector is installed on rack as shown in Figure 2-3. Secondary Flush Port Primary Flush Port Example of a 5 mm Instrument Figure 2-3.
  • Page 22: Figure 2-6. Reinstalling Connector

    4. Reinstall port connector in place, as show in Figure 2-6, by carefully sliding connector into place. Figure 2-6. Reinstalling Connector 5. Repeat Steps 1 to 4 until having all required port connectors ready. 6. Retrieve instruments with two ports on instrument housing. 7.
  • Page 23: Figure 2-8. Guiding Slider Into Holder Channel

    8. Ensure to open each instrument tip before processing load in washer/disinfector. 9. Locate instrument guiding slider and instrument holder channel. 10. While lowering instrument toward port connector, ensure to slide instrument guiding slider in instrument holder channel. Guiding Slider Holder NOTE: Drawings Channel...
  • Page 24: Figure 2-10. Securing Connection

    13. Once all instruments are connected, use one hand to secure each connection as follows: WARNING – INEFFECTIVE PROCESS HAZARD: • Place thumb below port connector; da Vinci instrument is not • Place remaining fingers on top of instrument housing; properly connected at the end of cycle, instrument has to be •...
  • Page 25: Da Vinci Instruments - Model Da Vinci X/Xi

    2.7.3 da Vinci Instruments — See Figures 2-12 to 2-15. Model da Vinci X/Xi 1. Locate port connectors on instrument holder. Each connector is installed on rack as shown in Figure 2-12. Secondary Flush Port Primary Flush Port Example of a 8 mm da Vinci X/Xi Instrument Figure 2-12.
  • Page 26: Figure 2-14. Connecting Instrument

    4. Ensure to open each instrument tip before processing load in washer/disinfector. 5. Once instrument is in front of port connector, carefully push on instrument housing to ensure both connector ends enter instrument housing flush ports. 6. Install all remaining instruments. Flush Ports Connector Ends...
  • Page 27: Figure 2-15. Installing Holding Cord

    7. Gently pull on device to ensure it remains connected. 8. Install holding cord on all instruments as shown in Figure 2-15. NOTE: Drawing Simplified For Clarity. Holding Cord Instrument Figure 2-15. Installing Holding Cord 2-17 Loading Instructions Operator Manual 11748808 FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 28 2-18 11748808 Operator Manual Loading Instructions FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 29: Operating Instructions

    OPERATING INSTRUCTIONS IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment can be found in S 1 of this manual. Do not use the equipment until you have become familiar with ECTION this information. 3.1 Robotic Cycle IMPORTANT: The RAS Rack is designed to be used only with the Operation...
  • Page 30: Cycle Description Chart

    3.2 Cycle Description See Table 3-1, Cycle Description Chart, for more information about Chart robotic cycle and parameters within cycle such as temperature, chemicals used, injection rate, etc. Use this chart as a reference while processing robotic cycle. 11748808 Operator Manual Operating Instructions FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 31 Operating Instructions Operator Manual 11748808 FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 32 11748808 Operator Manual Operating Instructions FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 33: Unloading Instructions

    NOTE: Some instruments might need to be slightly inclined to evacuate any remaining water inside ports. WARNING – SLIPPING HAZARD: To prevent slips, keep floors IMPORTANT: STERIS recommends unloading RAS Rack on an dry. Promptly clean up any appropriate surface or work table. spills or drippage. For spills or...
  • Page 34 11748808 Operator Manual Unloading Instructions FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 35: Routine Maintenance And Replacement Parts

    5.1 Preventive 1. Washer/disinfector regular service and maintenance MUST be Maintenance performed only by STERIS or a STERIS-trained technician. Any work performed by inexperienced or unqualified persons or the installation of unauthorized parts on washer/disinfector and rack could cause personal injury, invalidate the warranty or result in costly damage.
  • Page 36: Routine Maintenance

    Once a Month 5.2.3, R ECTION EPLACING LEXIBLE OSES ON NSTRUMENT OLDER -------------------------------------------------------------------------------------------------------------- Replace flexible hoses on rack instrument holder; or call STERIS* or Customer’s maintenance service to replace (see S 5.2.3, R Every Two Years ECTION EPLACING LEXIBLE OSES ON NSTRUMENT OLDER Verify holding cords on rack instrument holder.
  • Page 37: Figure 5-1. Removing Pin

    5.2.1 Inspecting and Cleaning See Figures 5-1 to 5-5. Rotary Spray Arm Assembly Once a week, the operator should inspect and clean chamber rotary rack spray arm assemblies as follows: 1. Remove pin securing rotary spray arm assembly to RAS Rack. Figure 5-1.
  • Page 38: Figure 5-3. Removing End Spray Jets

    3. Remove safety latch securing two end spray jets. 4. Remove two end spray jets and keep safely aside. Figure 5-3. Removing End Spray Jets 5. Use a fine wire (approximately wire gauge of a paper clip) to clean sediment from spray jets and end spray jets. If desired, cleaning needle can also be used (P117072-281).
  • Page 39 7. Reinstall end spray jets on rotary spray arm assembly using previously removed safety latches. 8. Once cleaning procedure is completed, reinstall rotary spray arm assembly to chamber ceiling using previously removed pin. 9. If required, once cleaning procedure is completed, have a qualified service technician...
  • Page 40: 5.2.2 Cleaning Rack Filter

    5.2.2 Cleaning Rack Filter See Figures 5-6 and 5-7. Once a week, the operator should let rack cool down and clean rack filter assembly as follows: 1. Unscrew filter from rack filter housing (one turn maximum); then, pull filter to remove it from filter housing. Filter Housing Filter Figure 5-6.
  • Page 41: Flexible Hoses Under Rack

    Hoses on Rack Instrument Every two years, or if flexible hose(s) are damaged, replace flexible Holder hoses as follows [or call STERIS or Customer’s maintenance service to replace]: NOTE: Routine Maintenance is not covered under warranty. •Flexible Hoses Under Rack NOTE: Flexible Hose Assembly Kit or Routine Maintenance Kit For Customer can be ordered from STERIS (see Table 5-2).
  • Page 42: Figure 5-10. Installing Hoses And Port Connectors

    4. Retrieve new hoses (see Table 5-2). 5. Pour some 70% isopropyl alcohol on end of hose to be reinstalled on irrigation port. Connect hose on rack. Repeat for remaining hoses. 6. Reinstall port connectors on hoses. 7. Reinstall port connectors on rack. 8.
  • Page 43: Flexible Hoses On Top Of Rack

    •Flexible Hoses On Top of Rack NOTE: Hose and Clip Assembly Kits or Routine Maintenance Kit For Customer can be ordered from STERIS (see Table 5-2). 1. Remove each hose from top connection. 2. Ensure to note type of hose clips in each position on rack instrument holder for later installation.
  • Page 44: Figure 5-13. Installing Hoses

    4. Retrieve new hoses (see Table 5-2). 5. Pour some 70% isopropyl alcohol on end of hose to be reinstalled on irrigation port. 6. While observing hose clip, ensure to connect hose on proper position of rack holder. 7. Repeat Steps 5 and 6 for remaining hoses. 8.
  • Page 45: Replacing Holding Cords On Rack Instrument Holder

    Cords on Rack Instrument Every two years, or if holding cord(s) are damaged, replace holding Holder cords as follows [or call STERIS* or Customer’s maintenance service to replace]: NOTE: Routine Maintenance is not covered under warranty. NOTE: Holding Cord or Routine Maintenance Kit For Customer can be ordered from STERIS (see Table 5-2).
  • Page 46: Figure 5-15. Removing Holding Cords

    3. Pull on holding cord to remove. Figure 5-15. Removing Holding Cords 5-12 11748808 Operator Manual Routine Maintenance and Replacement Parts FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 47 4. Retrieve new holding cords and O-rings (see Table 5-2). 5. Pour a small quantity of 70% isopropyl alcohol on both ends of each holding cord to be reinstalled on rack. 6. Install each holding cord in place. 7. While maintaining holding cord in place, install two new O-rings on other side of frame;...
  • Page 48: 5.3 Replacement Parts

    2. Specify model of your rack on your order (RAS Rack). 3. Send your order directly to STERIS. NOTE: Use only STERIS authorized parts on this equipment. Use of unauthorized parts will void the warranty. NOTE: See S 5.4, I...
  • Page 49 Table 5-2. Replacement Parts (Cont’d) Description Part Number CARTRIDGE, Filter 10038429 TIP, Filter 10038430 NUT, S/S, M5 x 0.8 mm, Filter 117-043-015 HOLDER, Tubing, Rack 10038727 CONNECTOR, Molded 10040737 ROD, Fixture, Filter Cartridge 10041442 O-RING, Silicone, 7/16" OD X 1/4" ID 10042072 SEAL, Handle, Filter 10052686...
  • Page 50: Illustrated Parts Breakdown

    Rack could cause personal injury, invalidate the warranty or result in costly damage. &'6#+. # &'6#+. $ 4'(  Figure 5-17. RAS 12 Rack (FD219) and RAS 12 Long Rack (FD220) – Part 1 of 2 5-16 11748808 Operator Manual Routine Maintenance and Replacement Parts...
  • Page 51 ITEM NUMBER ASS’Y 5-17 RAS 12 RACK (FD219) AND RAS 12 LONG RACK (FD220) – PART 1 OF 2 ..117043 015 NUT, S/S, M5 X 0.8 MM ..................117043 315 SCREW, PAN HEAD, S/S, M4 X 0.7 MM X 6 MM ..........
  • Page 52 &'6#+. % &'6#+. & & &'6#+. ' 4'(  Figure 5-18. RAS 12 Rack (FD219) and RAS 12 Long Rack (FD220) – Part 2 of 2 5-18 11748808 Operator Manual Routine Maintenance and Replacement Parts FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...
  • Page 53 DESCRIPTION ITEM NUMBER ASS’Y 5-18 RAS 12 RACK (FD219) AND RAS 12 LONG RACK (FD220) – PART 2 OF 2 ..117917 234 O-RING, SILICONE, 3/8" OD, X 3/16" ID............10077806 • KIT, RACK SERVICE MAINTENANCE (NOT SHOWN) ........10038727 HOLDER, TUBING, ENDOWRIST ..............
  • Page 54 5-20 11748808 Operator Manual Routine Maintenance and Replacement Parts FOR EXPORT ONLY, NOT FOR SALE OR USE IN USA...

This manual is also suitable for:

Ras 12 long rackFd220Fd219