Steris Reliance EPS Endoscope Processing System Operator's Manual page 111

Table of Contents

Advertisement

5-10
920011-649
10. Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) bar code reader. If NO is selected, display goes to Step 21.
11. Press ENTER touch pad to confirm setting. Display shows:
12. Operator ID function allows operator to scan a bar code to identify
processor operator. Operator ID is recorded on cycle printout.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) optional Operator ID function.
13. Press ENTER touch pad to confirm setting. Display shows:
14. Patient ID function allows operator to scan a bar code to identify
patient associated with endoscope to be processed. Patient ID
is recorded on cycle printout.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) optional Patient ID function.
15. Press ENTER touch pad to confirm setting. Display shows:
16. Device ID function allows operator to scan a bar code to identify
endoscope to be processed. Device ID is recorded on cycle
printout.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) optional Device ID function.
17. Press ENTER touch pad to confirm setting. Display shows:
18. Procedure ID function allows operator to scan a bar code to
identify procedure in which endoscope to be processed was
used. Procedure ID is recorded on cycle printout.
Press SCROLL arrows (up or down) to enable (YES) or disable
(NO) optional Procedure ID function.
Operator Manual
OPERATOR ID
YES
__ blinking
PATIENT ID
YES
__ blinking
DEVICE ID
YES
__ blinking
PROCEDURE ID
YES
__ blinking
Cycle and Control Value Programming