FAULT
Alarm sounds, and display shows:
FAULT 20
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 20
Alarm sounds, and display shows:
FAULT 21
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 21
Alarm sounds, and display shows:
FAULT 22
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 22
*Service charges may be incurred. Consult your warranty for details.
7-10
920011-649
1. Try to resume cycle by pressing START touch pad.
2. If situation reoccurs, press STOP touch pad twice to abort
3. When Abort sequence is completed, press UNLOCK
4. If door opens, close door and start cycle by pressing
5. If situation reoccurs, call STERIS.*
MM-DD-YY
=
1. Chamber temperature too high. After a D-SHORT or a
2. Air purge pre-filter needs changing, call STERIS.*
3. Air purge HEPA filter needs changing, call STERIS.*
4. Chamber RTD malfunction, call STERIS.*
MM-DD-YY
=
1. STOP touch pad was pressed during a cycle.
2. Press START touch pad to resume cycle or press STOP
MM-DD-YY
=
Operator Manual
POSSIBLE CAUSE AND CORRECTION
DOOR SYSTEM MALFUNCTION
Door may not be closed properly. Proceed as follows:
cycle.
toggle switch.
START touch pad.
SUMP RTD OVERHEAT
Possible causes:
D-LONG cycle, leave door open and allow unit and
support to cool before starting a new cycle.
STOP PRESSED
touch
pad to abort cycle.
Troubleshooting