FAULT
Alarm sounds, and display shows:
FAULT 40
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 40
Alarm sounds, and display shows:
FAULT 41
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 41
Alarm sounds, and display shows:
FAULT 42
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM
---------------------------------------
=
FAULT 42
*Service charges may be incurred. Consult your warranty for details.
7-18
920011-649
MM-DD-YY
=
MM-DD-YY
=
1. Verify hot tap water (HTW) supply temperature is as per
2. Try to resume cycle by pressing START touch pad. If
MM-DD-YY
=
Operator Manual
POSSIBLE CAUSE AND CORRECTION
LEAK TEST PORT 1: PRESSURE SENSOR DEFECTIVE
Press ALARM REPLY touch pad to silence audible alarm.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
LEAK TEST PORT 2: PRESSURE SENSOR DEFECTIVE
Press ALARM REPLY touch pad to silence audible alarm.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
FILLING WATER TEMPERATURE TOO HIGH
equipment
drawing
temperature, as read by sump RTD, can be verified by
pressing SINGLE or DOUBLE touch pad twice during
filling phase of a cycle.
situation reoccurs, call STERIS.*
(920-011-656).
Filling
Troubleshooting
water